K253649

Philips Medical Systems Technologies , Ltd. · cleared 2026-03-27 · product code JAK · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.4)
The Spectral CT Verida Family is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.The system incorporates both conventional iterative reconstruction (IR) and artificial intelligence (AI)-based reconstruction functionality. Spectral Precise Image (SPI) is an Al-based deep learning reconstruction feature intended to optimize image quality by reducing noise and enhancing image appearance in Head, Whole Body, Cardiac, and Vascular X-ray Computed Tomography applications.
AlgorithmAI-based deep learning reconstruction
source quote (p.8)
The system incorporates both conventional iterative reconstruction (IR) and artificial intelligence (AI)-based reconstruction functionality. Spectral Precise Image (SPI) is an Al-based deep learning reconstruction feature intended to optimize image quality by reducing noise and enhancing image appearance in Head, Whole Body, Cardiac, and Vascular X-ray Computed Tomography applications.
Adaptive (vs locked)No
source quote (p.9)
SPI is derived from the previously cleared Precise Image framework and does not alter the fundamental image-formation process or introduce new clinical decision-making functionality.
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.36)
The cybersecurity testing for the Spectral CT Verida Family system was conducted in accordance with FDA-recognized consensus standards and specifically aligns with the following FDA guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (FDA, June 27, 2025) Additional relevant FDA software and risk management guidance documents.

Validation studies (2)

Bench

n=35 scans

endpoints: CT number linearity; image noise; uniformity; contrast-to-noise ratio (CNR); spatial resolution (MTF and SSP); low contrast resolution (visual); image artifacts; Monochromatic energy (MonoE, HU); virtual non-contrast (VNC, HU); iodine density (mg/ml); iodine no water (mg/ml); iodine removed (HU); contrast-enhanced structures (HU); calcium suppression (HU); Z effective; electron density (%EDW); uric acid and uric acid removed (HU)

Retrospective clinical

n=147 scans · 2 site(s)

endpoints: Non-inferiority of SPI vs. iDose⁴ for image quality and diagnostic confidence in both conventional and spectral images.; Comparison of SPI spectral images to iDose conventional images; performance in chest high-resolution scans; success rate; reader agreement

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

75
recalls in product code, 24mo
192
MAUDE reports in code, 12mo
-22%
vs code's own 3-yr baseline
9
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98738

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98588

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:97699

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98206

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97010

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (, initiated 2025-05-23): "Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96967

  • …and 3 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2026-03-26

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    recall event 98738 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2026-03-26

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    recall event 98738 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2026-03-26

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    recall event 98738 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2026-03-26

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    recall event 98738 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.