RT Elements (4.5); (Elements) Multiple Brain Mets SRS; (Elements) Cranial SRS; (Elements) Spine SRS; (Elements) Cranial SRS w/ Cones; (Elements) RT Planning Platform; (Elements) Dose Review; (Elements) Retreatment Review; Elements Segmentation [Cranial , Basal Ganglia, Head & Neck, Pelvic, Spine, Thoracic & Spine, Extracranial] RT; Elements AI Tumor Segmentation RT; Elements SmartBrush [Angio, Spine] RT; Elements Object Management RT

K250440

Brainlab AG · cleared 2025-06-17 · product code MUJ · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.7)
The device is composed by software only and can be controlled via mouse and keyboard. The software is displayed on a computer screen. The Cranial Tumor Segmentation feature uses Al/Ml.
AlgorithmAI/ML algorithm for cranial tumor segmentation, trained on MRI image data.
source quote (p.6)
The Cranial Tumor Segmentation feature uses Al/Ml. The algorithm was trained on MRI image data with contrast-enhancing tumors from multiple clinical sites, including a wide variety of scanner models and patient characteristics.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=412 patients

endpoints: Dice coefficient; Recall; Precision

Reported performance (3 observations)

sensitivity0.85CI [0.83; 0.87]
source quote (p.15)
Table 1 Summary of test statistics
diceas written: “Dice0.75CI [0.74; 0.76]
source quote (p.15)
Table 1 Summary of test statistics
ppvas written: “Precision0.86CI [0.85; 0.88]
source quote (p.15)
Table 1 Summary of test statistics

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

35
recalls in product code, 24mo
17
MAUDE reports in code, 12mo
-50%
vs code's own 3-yr baseline
2
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-08-05): "Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet Firs" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97049

  • recall_reason_pattern

    Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-07-17): "Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ," Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97309

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K250440