Neuro Insight V1.0

K243681

Olea Medical S.A.S. · cleared 2025-07-23 · product code LLZ · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.4)
Neuro Insight V1.0 is an image processing solution. It is intended to assist appropriately trained medical professionals in their analysis workflow on neurological MRI images.
AlgorithmThis brain extraction algorithm uses a 3D U-Net architecture, consisting of 6 layers (16, 32, 64, 128, 256, 512), and was optimized using both DICE and Tversky metrics.
source quote (p.8)
This brain extraction algorithm uses a 3D U-Net architecture, consisting of 6 layers (16, 32, 64, 128, 256, 512), and was optimized using both DICE and Tversky metrics.
Adaptive (vs locked)No
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=100 cases

endpoints: DICE coefficient

Reported performance (1 observation)

diceas written: “DICE coefficient0.97CI 0.907 to 0.988
source quote (p.9)
The achieved average DICE coefficient of 0.97, ranging from 0.907 to 0.988, exceeds the predetermined acceptance threshold of 0.95.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

48
recalls in product code, 24mo
295
MAUDE reports in code, 12mo
+683%
vs code's own 3-yr baseline
26
drift signals on this device
  • adverse_event_inflection

    MAUDE adverse-event reports for product code LLZ: 295 in the 12 months ending 2026-06, vs a 37.7/12mo average over the prior 3 windows (+683%). Code-level count — reports are not attributed to this specific device.

    first seen 2026-07-08 · openFDA /device/event.json count=date_received product_code=LLZ

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems SCS, initiated 2026-05-08): "GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:99042

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2026-03-05): "Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98520

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems, LLC, initiated 2026-01-30): "There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98428

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Focalyx Technologies, LLC., initiated 2025-12-23): "Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, wh" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98137

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (DICOM Grid, Inc., initiated 2025-11-18): "Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97993

  • …and 20 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (OLEA MEDICAL 93 AVANUE DU SORBIER, ZONE ATHELIA IV LA CIOTAT France) — same firm and product code, not necessarily this device · initiated 2023-08-10

    When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.

    recall event 93049 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K243681