BoneMetrics (US)

K241593

Gleamer SAS · cleared 2025-02-05 · product code QIH · Radiology

FDA 510(k) summary for BoneMetrics (US) (K241593)

Constat organizes this clearance into predicate context, validation evidence, reported performance, postmarket signals, and reimbursement links. Each extracted claim stays tied to the public FDA source; missing fields remain explicitly unstated.

Validation studies
1
Cited predicates
1
Performance observations
0
Product code
QIH

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.5)
“BoneMetrics US is a fully automated radiological image processing software device intended to aid users in the measurement of Cobb angles during the review of frontal spine radiographs. The software device is to be used by healthcare professionals trained in radiology. Software application operated on OTS software.”
AlgorithmAI algorithm identifies keypoints corresponding to the corners of all the vertebras and calculates Cobb angles. It uses convolutional neural networks for classification and landmarking, and classical methods for computing angles.
source quote (p.6)
“Once received by BoneMetrics US, the radiographs are automatically processed by the AI algorithm without requiring any user inputs. The algorithm identifies the keypoints corresponding to the corners of all the vertebras that are seen on the images and calculates all possible angles between vertebras. Only Cobb Angles that are above 7º are retained. Convolutional neural networks for: classification, landmarking. Classical methods for computing: angles”
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedNo

Validation studies (1)

Retrospective clinical

n=345 images

endpoints: Cobb angle with the largest curvature; Minor Cobb angle

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

1
recalls in product code, 24mo
4
MAUDE reports in code, 12mo
—
vs code's own 3-yr baseline
1
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code QIH (PIE Medical Imaging B.V., initiated 2026-05-07): "User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical d" Recalling firm is another firm in the same product code.

    first seen 2026-08-26 · recall res_event_number:98930

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-08-26.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Related FDA clearances

These are recent clearances in the same FDA product-code category. Shared classification does not by itself make a device an appropriate predicate.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K241593