AiMIFY (1.x)

K240290

Subtle Medical, Inc. · cleared 2024-08-21 · product code LLZ · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.6)
The AiMIFY device is a software as a medical device consisting of a machine learning software algorithm that enhances images taken by MRI scanners.
Algorithmmachine learning software algorithm, convolutional neural network based algorithm, convolutional neural network based filtering, UNet style neural network
source quote (p.6)
The AiMIFY device is a software as a medical device consisting of a machine learning software algorithm that enhances images taken by MRI scanners. AiMIFY image processing software uses a convolutional neural network based algorithm to enhance the AiMIFY-contrast images from pre-contrast and standard-dose post-contrast images. AiMIFY implements an image enhancement algorithm using convolutional neural network based filtering. The model performs non-linear scaling of the contrast uptake between the pre- and post-contrast images to improve the visualization of the areas in the brain that have absorbed the contrast material. AIMIFY processes the input image in a 2.5D fashion using a UNet style neural network.
Adaptive (vs locked)Yes
source quote (p.8)
AiMIFY has a Predetermined Change Control Plan (PCCP) for planned changes to improve the generalizability of the AiMIFY Artificial Intelligence/Machine Learning (AI/ML) model, to reduce processing time, and improve perceived image quality.
PCCPYes
source quote (p.1)
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP).
Cybersecurity addressedYes
source quote (p.6)
The AiMIFY device itself is not networked and therefore does not increase the cybersecurity risk of its users. Users are provided cybersecurity recommendations in labeling.

Validation studies (2)

Bench

n=95 cases

endpoints: CNR improvement; LBR improvement; CEP improvement

Reader study (MRMC)

n=95 cases

endpoints: visibility of lesion features (lesion contrast enhancement, border delineation, and internal morphology); visibility of perceived image quality and image artifacts; Vessel Conspicuity; False Lesion Analysis; Lesion Size; Radiomics

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

48
recalls in product code, 24mo
295
MAUDE reports in code, 12mo
+683%
vs code's own 3-yr baseline
26
drift signals on this device
  • adverse_event_inflection

    MAUDE adverse-event reports for product code LLZ: 295 in the 12 months ending 2026-06, vs a 37.7/12mo average over the prior 3 windows (+683%). Code-level count — reports are not attributed to this specific device.

    first seen 2026-07-08 · openFDA /device/event.json count=date_received product_code=LLZ

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems SCS, initiated 2026-05-08): "GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:99042

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2026-03-05): "Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98520

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems, LLC, initiated 2026-01-30): "There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98428

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Focalyx Technologies, LLC., initiated 2025-12-23): "Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, wh" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98137

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (DICOM Grid, Inc., initiated 2025-11-18): "Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97993

  • …and 20 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.