K233582

iSchema View Inc. · cleared 2024-04-22 · product code LLZ · Radiology

FDA 510(k) summary for Rapid (K233582)

Constat organizes this clearance into predicate context, validation evidence, reported performance, postmarket signals, and reimbursement links. Each extracted claim stays tied to the public FDA source; missing fields remain explicitly unstated.

Validation studies
1
Cited predicates
5
Performance observations
0
Product code
LLZ

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
“Rapid is an image processing software package to be used by trained professionals, including but not limited to physicians (medical analysis and decision making) and medical technicians (administrative case processing). The software runs on a standard off-the-shelf computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing, and analysis of images. Data and images are acquired through DICOM compliant imaging devices. Rapid is indicated for use in Adults only.”
AlgorithmMixed Traditional and AI/ML
source quote (p.12)
“Mixed Traditional and AI/ML”
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.10)
“Rapid has been designed and validated to the FDA Cybersecurity Guidance and IEC 81001-5-1.”

Validation studies (1)

Bench

sample size not stated

endpoints: accurate representation of key processing parameters

standards: EN ISO 14971:2019, IEC 62304:2016, IEC 62366:2015, NEMA PS 3.1 - 3.20, IEC 81001-5-1

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

25
recalls in product code, 24mo
397
MAUDE reports in code, 12mo
+963%
vs code's own 3-yr baseline
32
drift signals on this device
  • predicate_cohort_contagion

    K233582 shares predicate K121447 with K233512 (Rapid (6.0)), whose product code QIH recorded 1 recall event(s) in the trailing 24 months.

    first seen 2026-09-02 · predicate:K121447 sibling:K233512

  • predicate_cohort_contagion

    K233582 shares predicate K182130 with K233209 (uOmnispace.CT), whose product code QIH recorded 1 recall event(s) in the trailing 24 months.

    first seen 2026-09-02 · predicate:K182130 sibling:K233209

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Merge Healthcare, Inc., initiated 2026-07-22): "Software issue may cause measurements to be displayed inaccurately." Recalling firm is another firm in the same product code.

    first seen 2026-09-02 · recall res_event_number:99520

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (FUJIFILM Healthcare Americas Corporation, initiated 2026-07-22): "Software issue may cause the ruler measurement to be inaccurate." Recalling firm is another firm in the same product code.

    first seen 2026-09-02 · recall res_event_number:99529

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Merge Healthcare, Inc., initiated 2026-06-30): "Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient st" Recalling firm is another firm in the same product code.

    first seen 2026-09-02 · recall res_event_number:99391

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Boston Scientific Corporation, initiated 2026-06-19): "Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an inpu" Recalling firm is another firm in the same product code.

    first seen 2026-09-02 · recall res_event_number:99236

  • …and 26 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-09-02.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.