Swoop® Portable MR Imaging System®

K232760

Hyperfine, Inc. · cleared 2023-10-06 · product code LNH · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.5)
The Swoop system is portable, ultra-low field MRI device that enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. ... The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, and FLAIR sequences, specifically in terms of reductions in image noise and blurring.
AlgorithmThe image reconstruction algorithm utilizes deep learning to provide improved image quality. Specific algorithms include Advanced Gridding for T1W, T2W, and FLAIR sequences, Fast Iterative Shrinkage Thresholding Algorithm (FISTA) for DWI, and Advanced Denoising for T1W, T2W, FLAIR, and DWI.
source quote (p.5)
The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, and FLAIR sequences, specifically in terms of reductions in image noise and blurring. ... Advanced Gridding ... Fast Iterative Shrinkage Thresholding Algorithm (FISTA) ... Advanced Denoising (applies to T1W, T2W, FLAIR and DWI)
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.8)
Testing to verify cybersecurity controls and management. ... Cybersecurity as recommended in FDA guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"

Validation studies (1)

Bench

sample size not stated

endpoints: image quality criteria

standards: IEC 62304:2015, FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", NEMA MS 1-2008 (R2020), NEMA MS 3-2008 (R2020), NEMA MS 9-2008 (R2020), NEMA MS 12-2016, American College of Radiology (ACR) Phantom Test Guidance for Use of the Large MRI Phantom for the ACR MRI Accreditation Program, American College of Radiology standards for named sequences, ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2010, FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling", ISO 17664:2017, ASTM F3208-17, ANSI/AAMI ES 60601-1:2005/(R)2012, IEC 60601-1-2:2014, IEC 60601-1-6:2013, NEMA MS 8-2016, FDA guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

100
recalls in product code, 24mo
510
MAUDE reports in code, 12mo
+5%
vs code's own 3-yr baseline
6
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253489 (decision 2025-12-12) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253489

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K250236 (decision 2025-05-30) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System (V2)") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K250236

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251276 (decision 2025-05-21) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251276

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K240944 (decision 2024-07-16) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K240944

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98779

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98111

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K232760