ACUSON Sequoia Diagnostic Ultrasound System, ACUSON Sequoia Select Diagnostic Ultrasound System, ACUSON Origin Diagnostic Ultrasound System

K232145

Siemens Medical Solutions USA, Inc. · cleared 2023-10-30 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.6)
The ACUSON Sequoia, Sequoia Select, and Origin Diagnostic Ultrasound Systems are multi-purpose, mobile, software-controlled, diagnostic ultrasound systems with an on-screen display of thermal and mechanical indices related to potential bio-effect mechanisms. The ultrasound system function is to transmit and receive ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M Mode, Doppler Tissue Mode, Amplitude Doppler Mode, a combination of modes, Panoramic Imaging, Contrast agent Imaging, Virtual Touch Strain Imaging (except Origin), Virtual Touch – pSWE Imaging, Virtual Touch – SWE Imaging, Custom Tissue Imaging, 3D/4D Volume Imaging or Harmonic Imaging on a Display and provide cardiac anatomical and quantitative function software applications. The ACUSON Sequoia, Sequoia Select, and Origin will offer an optional cardiac imaging and analysis package with new transducers and software applications with Artificial Intelligent algorithms.
AlgorithmArtificial Intelligent algorithms; Machine Learning that are specifically intended to support cardiac imaging for blood flow, valves and anatomical features
source quote (p.8)
The ACUSON Sequoia, Sequoia Select, and Origin will offer an optional cardiac imaging and analysis package with new transducers and software applications with Artificial Intelligent algorithms. The ACUSON Sequoia, Sequoia Select, and Origin will offer four (4) new software applications (Al Measure, Al Assist, 2D Heart^l, 4D Heart^l) containing Machine Learning that are specifically intended to support cardiac imaging for blood flow, valves and anatomical features and modifies the Indications for Use.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (4)

Retrospective clinical

n=12 patients

endpoints: algorithm success

Retrospective clinical

n=45 patients · 5 site(s)

endpoints: Pearson correlation coefficient; Bland Altman bias

Retrospective clinical

n=32 patients · 5 site(s)

endpoints: Pearson correlation coefficient; Bland Altman bias

Retrospective clinical

n=32 patients · 5 site(s)

endpoints: algorithm success

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
7
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:95254

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251481 (decision 2025-08-20) from Siemens Medical Solutions USA, Inc. for a matching device line ("ACUSON Sequoia Diagnostic Ultrasound System; ACUSON Sequoia Select Diagnostic Ultrasound System; ACUSON Origin Diagnostic Ultrasound System; ACUSON Origin ICE Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251481

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K242523 (decision 2024-11-19) from Siemens Medical Solutions USA, Inc. for a matching device line ("ACUSON Sequoia Diagnostic Ultrasound System; ACUSON Sequoia Select Diagnostic Ultrasound System; ACUSON Origin Diagnostic Ultrasound System; ACUSON Origin ICE Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K242523

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K240704 (decision 2024-07-22) from Siemens Medical Solutions, USA, Inc. for a matching device line ("ACUSON Sequoia Diagnostic Ultrasound System, ACUSON Sequoia Select Diagnostic Ultrasound System, ACUSON Origin Diagnostic Ultrasound System, ACUSON Origin ICE Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K240704

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97726

  • …and 1 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc.) — same firm and product code, not necessarily this device · initiated 2024-08-15

    If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

    recall event 95254 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc.) — same firm and product code, not necessarily this device · initiated 2024-08-15

    If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

    recall event 95254 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc.) — same firm and product code, not necessarily this device · initiated 2024-08-15

    If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems will perform incorrect unit conversion of liters to milliliters using multiplier 100 instead of 1000, resulting values 10x smaller than actual, which could contribute misdiagnosis or negatively influence patient management

    recall event 95254 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc.) — same firm and product code, not necessarily this device · initiated 2024-02-23

    On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

    recall event 94211 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.