Global Hypoperfusion Index (GHI) Algorithm

K231038

Edwards Lifesciences, LLC · cleared 2023-07-26 · product code QNL · Cardiovascular

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
The Global Hypoperfusion Index (GHI) algorithm provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability.
Algorithmmachine-learning model
source quote (p.6)
Predictive algorithm using a machine-learning model
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

sample size not stated

endpoints: safety; performance

standards: FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005), ANSI/AAMI/IEC 62304:2006/A1:2016, Medical Device Software – Software Life Cycle Processes, FDA Guidance General Principles of Software Validation (issued January 11, 2022)

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Cardiovascular panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K231038