Vantage Galan 3T, MRT-3020, V9.0 with AiCE Reconstruction Processing Unit for MR

K230355

Canon Medical Systems Corporation · cleared 2023-08-30 · product code LNH · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.6)
The Vantage Galan (Model MRT-3020) is a 3 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K222387. This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. Iterative Motion Correction (IMC): IMC has been updated to utilize Deep Learning based methods in addition to traditional model-based methods.
AlgorithmDeep Learning based methods, Deep Learning reconstruction, Machine Learning based method
source quote (p.6)
Iterative Motion Correction (IMC): IMC has been updated to utilize Deep Learning based methods in addition to traditional model-based methods. Free Breathing Dynamic: Free Breathing Dynamic is contrast enhanced dynamic imaging technique. Utilizing Deep Learning reconstruction, Free Breathing Dynamic enables a single continuous scan to aid patients who have difficulty holding their breath. Precise IQ Engine (PIQE): PIQE is Deep Learning based technique that generates higher in-plane matrix images from low matrix images while mitigating the ringing artifact. NeuroLine+: NeuroLine+ has been updated to utilize Machine Learning based method.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedNo

Validation studies (17)

Bench

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

sample size not stated

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Reader study (MRMC)

n=17 patients

endpoints: ringing; sharpness; SNR; overall IQ; feature conspicuity

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Retrospective clinical

n=13 patients

endpoints: successful scan alignment (offset and angle within acceptable error); accuracy of the auto-detected angle or center position of the target planes

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

n=12 cases

endpoints: peak SNR; structural similarity (SSIM)

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Reader study (MRMC)

n=21 patients

endpoints: SNR; tissue contrast; image sharpness; diagnostic confidence

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Reader study (MRMC)

n=34 images

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Bench

n=18 patients

endpoints: success rate of automatically proposed phases

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Reader study (MRMC)

n=29 patients

endpoints: overall SNR; overall IQ; feature conspicuity; diagnostic confidence

standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, IEC60601-1-2 (2020), IEC60601-1-6 (2010), Amd.1 (2013), Amd.2 (2020), IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015), IEC60825-1 (2014), IEC62304 (2006), Amd.1 (2015), IEC62366-1 (2020), NEMA MS 1 (2008), NEMA MS 2 (2008), NEMA MS 3 (2008), NEMA MS 4 (2010), NEMA MS 5 (2010)

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

100
recalls in product code, 24mo
510
MAUDE reports in code, 12mo
+5%
vs code's own 3-yr baseline
4
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K243335 (decision 2025-01-07) from Canon Medical Systems Corporation for a matching device line ("Vantage Galan 3T, MRT-3020, V10.0 with AiCE Reconstruction Processing Unit for MR") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K243335

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K241496 (decision 2024-08-20) from Canon Medical Systems Corporation for a matching device line ("Vantage Galan 3T, MRT-3020, V10.0 with AiCE Reconstruction Processing Unit for MR") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K241496

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98779

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98111

Recalls attributed to this device

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K230355