BriefCase

K230020

Aidoc Medical, Ltd. · cleared 2023-02-01 · product code QFM · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.4)
BriefCase is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.
Algorithmartificial intelligence, deep-learning algorithms
source quote (p.5)
Both devices are artificial intelligence, deep-learning algorithms incorporated in software packages for use with DICOM compliant CT scanners, PACS, and radiology workstations.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=308 cases · 5 site(s)

endpoints: area under the curve (AUC); BriefCase time-to-notification; Positive Predictive Value (PPV); Negative Predictive Value (NPV); Positive Likelihood Ratio (PLR); Negative Likelihood Ratio (NLR)

Reported performance (8 observations)

sensitivity0.9808CI 93.23%, 99.77%
source quote (p.8)
Sensitivity was 98.08% (95% CI: 93.23%, 99.77%)
specificity0.9314CI 88.75%, 96.20%
source quote (p.8)
Specificity was 93.14% (95% CI: 88.75%, 96.20%).
aurocas written: “auc0.98CI 0.966, 0.994
source quote (p.8)
AUC was 0.98 (95% CI: 0.966, 0.994)
npvas written: “NPV0.998CI 99.1% - 99.9%
source quote (p.9)
NPV was 99.8% (95% CI: 99.1% - 99.9%)
ppvas written: “PPV0.614CI 48.9% - 72.5%
source quote (p.9)
PPV was 61.4% (95% CI: 48.9% - 72.5%).
ppvas written: “PLR14.2912CI 8.614 - 23.710
source quote (p.9)
PLR was 14.2912 (95% CI: 8.614 - 23.710)
npvas written: “NLR0.0206CI 0.005 - 0.082
source quote (p.9)
NLR was 0.0206 (95% CI: 0.005 - 0.082).
time_to_resultas written: “Time-to-notification70.1CI 64.9-75.4
source quote (p.8)
The BriefCase mean time-to-notification for the subject RibFx triage was 70.1 seconds (95% CI: 64.9-75.4).

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
15
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253578 (decision 2026-02-26) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253578

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251195

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K252970 (decision 2026-01-07) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-triage CT Body") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K252970

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253265 (decision 2025-11-06) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253265

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251406 (decision 2025-05-30) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251406

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K250248 (decision 2025-02-14) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K250248

  • …and 9 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K230020