RadioLens v1.0

K222174

Synapsica · cleared 2023-03-28 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.4)
RadioLens v1.0 includes an optional SpindleX module. The SpindleX is an artificial intelligence (AI) module that works on digital stress x-rays of the spine and assists with clinical interpretation
AlgorithmArtificial intelligence (AI) module
source quote (p.4)
The SpindleX is an artificial intelligence (AI) module that works on digital stress x-rays of the spine and assists with clinical interpretation
Adaptive (vs locked)No
source quote (p.5)
These pre-marked points were treated as ground truth for training of models
PCCPNo
Cybersecurity addressedYes
source quote (p.7)
Content of Premarket Submission for Management of Cybersecurity in Medical Devices.

Validation studies (1)

Reader study (MRMC)

n=600 images · 200 site(s)

endpoints: sensitivity; specificity; agreement (ICC scores)

standards: ANSI AAMI IEC 62304:2006/A1:2016, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Content of Premarket Submission for Management of Cybersecurity in Medical Devices

Reported performance (20 observations)

sensitivityas written: “Sensitivity for identification of body part among cervical and lumbar scans100
source quote (p.8)
For identification of body part among cervical and lumbar scans, the model had 100% sensitivity for the scans used in the validation study.
sensitivityas written: “Sensitivity for identification of patient position among flexion96.8
source quote (p.8)
For identification of patient position among flexion, extension and neutral the model had 96.8%, 92.0%, and 90.5% sensitivities respectively.
sensitivityas written: “Sensitivity for identification of patient position among extension92
source quote (p.8)
For identification of patient position among flexion, extension and neutral the model had 96.8%, 92.0%, and 90.5% sensitivities respectively.
sensitivityas written: “Sensitivity for identification of patient position among neutral90.5
source quote (p.8)
For identification of patient position among flexion, extension and neutral the model had 96.8%, 92.0%, and 90.5% sensitivities respectively.
agreement_kappaas written: “Rads & Model ICC - Stress Lines (degrees) - Cervical0.953CI E
source quote (p.8)
Stress Lines (degrees) 0.953 (E)
agreement_kappaas written: “Rads & Model ICC - Canal Diameter (mm) - Cervical0.955CI E
source quote (p.8)
Canal Diameter (mm) 0.955 (E)
agreement_kappaas written: “Rads & Model ICC - Vertebral Offset Neutral (mm) - Cervical0.707CI M
source quote (p.8)
Vertebral Offset Neutral (mm) 0.707 (M)
agreement_kappaas written: “Rads & Model ICC - Angular MSI (degrees) - Cervical0.781CI G
source quote (p.8)
Angular MSI (degrees) 0.781 (G)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 5th Edition (mm) - Cervical0.717CI M
source quote (p.8)
Translational Motion 5th Edition (mm) 0.717 (M)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 6th Edition - Flexion (mm) - Cervical0.802CI G
source quote (p.8)
Translational Motion 6th Edition - Flexion (mm) 0.802 (G)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 6th Edition - Extension (mm) - Cervical0.716CI M
source quote (p.8)
Translational Motion 6th Edition - Extension (mm) 0.716 (M)
agreement_kappaas written: “Rads & Model ICC - Stress Lines (degrees) - Lumbar0.968CI E
source quote (p.8)
Stress Lines (degrees) 0.968 (E)
agreement_kappaas written: “Rads & Model ICC - Canal Diameter (mm) - Lumbar0.784CI G
source quote (p.9)
Canal Diameter (mm) 0.784 (G)
agreement_kappaas written: “Rads & Model ICC - Ferguson's Angle (degrees) - Lumbar0.939CI E
source quote (p.9)
Ferguson's Angle (degrees) 0.939 (E)
agreement_kappaas written: “Rads & Model ICC - Integrity of 3rd Lumbar Vertebra (mm) - Lumbar0.996CI E
source quote (p.9)
Integrity of 3rd Lumbar Vertebra (mm) 0.996 (E)
agreement_kappaas written: “Rads & Model ICC - Vertebral Offset Neutral (mm) - Lumbar0.777CI G
source quote (p.9)
Vertebral Offset Neutral (mm) 0.777 (G)
agreement_kappaas written: “Rads & Model ICC - Angular MSI (degrees) - Lumbar0.771CI G
source quote (p.9)
Angular MSI (degrees) 0.771 (G)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 5th Edition (mm) - Lumbar0.656CI M
source quote (p.9)
Translational Motion 5th Edition (mm) 0.656 (M)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 6th Edition - Flexion (mm) - Lumbar0.774CI G
source quote (p.9)
Translational Motion 6th Edition - Flexion (mm) 0.774 (G)
agreement_kappaas written: “Rads & Model ICC - Translational Motion 6th Edition - Extension (mm) - Lumbar0.769CI G
source quote (p.9)
Translational Motion 6th Edition - Extension (mm) 0.769 (G)

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).