K221314

Aidoc Medical, Ltd. · cleared 2022-06-03 · product code QAS · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of M1 Large Vessel Occlusion (M1-LVO) pathologies.
Algorithmartificial intelligence, deep-learning algorithm
source quote (p.3)
BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=383 patients

endpoints: Sensitivity; Specificity; Time-to-notification; NPV; PPV

Reported performance (6 observations)

sensitivity0.888CI 81.9%, 93.8%
source quote (p.7)
Sensitivity was 88.8% (95% CI: 81.9%, 93.8%)
specificity0.872CI 82.5%, 91.1%
source quote (p.7)
specificity was 87.2% (95% CI: 82.5%, 91.1%).
npvas written: “NPV0.995CI 99.3%-99.8%
source quote (p.8)
NPV was 99.5% (95% CI: 99.3%-99.8%)
ppvas written: “PPV0.201CI 13.2%-24.2%
source quote (p.8)
PPV was 20.1% (95% CI: 13.2%-24.2%).
time_to_resultas written: “Time-to-notification of BriefCase M1-LVO3.8CI 3.6-4.0
source quote (p.8)
The BriefCase mean time-to-notification for M1-LVO was 3.8 minutes (95% CI: 3.6-4.0).
time_to_resultas written: “Time-to-notification of BriefCase ICH K1806474.46CI 4.10-4.83
source quote (p.8)
The time-to-notification for the BriefCase for ICH triage was 4.46 minutes (95% CL: 4.10-4.83).

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
21
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253578 (decision 2026-02-26) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253578

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251195

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K252970 (decision 2026-01-07) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-triage CT Body") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K252970

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253265 (decision 2025-11-06) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253265

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251406 (decision 2025-05-30) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251406

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K250248 (decision 2025-02-14) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K250248

  • …and 15 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K221314