Vivid iq

K221148

GE Medical Systems Ultrasound & Primary Care Diagnostics LLC · cleared 2022-07-18 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.5)
The proposed Vivid iq system is a general-purpose, Track 3, diagnostic ultrasound device, primarily intended for cardiovascular diagnostic use and shared service imaging. It is an ultrasound imaging & analysis system, consisting of a compact console with control panel including a track pad, color LCD Touch Panel that includes an on-screen alfa-numeric keyboard. The system also has an optional height-adjustable cart for comfortable standing and sitting positions. An extended battery is integrated within the Vivid iq cart and provides additional power for longer scanning time. Added Easy Auto EF -based on Auto EF 3.0 (includes AI Auto ROI algorithm)
AlgorithmAI algorithm (average dice score)
source quote (p.7)
The accuracy of the AI algorithm (average dice score) as tested on datasets from different countries, is 92% or higher; as tested on datasets from different scanning views, is 91% or higher; as tested on dataset from different left ventricle volumes, is 92% or higher.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=45 patients

endpoints: accuracy (average dice score)

Reported performance (1 observation)

accuracyas written: “accuracy (average dice score)0.92
source quote (p.7)
The accuracy of the AI algorithm (average dice score) as tested on datasets from different countries, is 92% or higher; as tested on datasets from different scanning views, is 91% or higher; as tested on dataset from different left ventricle volumes, is 92% or higher.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
7
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:97726

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251169 (decision 2025-07-10) from GE Medical Systems Ultrasound and Primary Care Diagnostics for a matching device line ("Vivid Pioneer") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251169

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K243628 (decision 2025-02-11) from GE Medical Systems Ultrasound and Primary Care Diagnostics for a matching device line ("Vivid T9/Vivid T8") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K243628

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K243620 (decision 2025-02-11) from GE Medical Systems Ultrasound and Primary Care Diagnostics for a matching device line ("Vivid iq") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K243620

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • …and 1 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-03-06

    The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.

    recall event 96538 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K221148