MAGNETOM Avanto Fit

K220151

Siemens Medical Solutions USA, Inc. · cleared 2022-04-01 · product code LNH · Radiology

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.3)
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.
AlgorithmImage-based motion correction, automatic fiducial detection, shimming procedure, speed-optimized reference scan for GRAPPA and SMS kernel calibration.
source quote (p.6)
TSE MoCo: TSE MoCo is an image-based motion correction in the average-dimension for the TSE pulse sequence type. Automatic fiducial detection: MR Breast Biopsy is improved with an automatic fiducial detection. AbsoluteShim: The AbsoluteShim mode is a shimming procedure based on a 3-echo gradient echo protocol. Fast GRE RefScan: A speed-optimized reference scan for GRAPPA and SMS kernel calibration for echo planar imaging pulse sequence types.
Adaptive (vs locked)No
PCCPNo
Cybersecurity addressedNo

Validation studies (4)

Bench

sample size not stated

endpoints: Image quality assessments

Bench

sample size not stated

Bench

sample size not stated

Bench

sample size not stated

standards: AAMI / ANSI ES60601-1, IEC 60601-2-33

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

100
recalls in product code, 24mo
510
MAUDE reports in code, 12mo
+5%
vs code's own 3-yr baseline
3
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K232494 (decision 2023-11-14) from Siemens Medical Solutions USA, Inc. for a matching device line ("MAGNETOM Avanto fit; MAGNETOM Skyra fit") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K232494

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98779

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98111

Recalls attributed to this device

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc) — same firm and product code, not necessarily this device · initiated 2025-10-01

    There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    recall event 97845 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc) — same firm and product code, not necessarily this device · initiated 2025-08-28

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    recall event 97530 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc) — same firm and product code, not necessarily this device · initiated 2025-08-28

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    recall event 97530 (openFDA)

  • Recalling firm matches this device's applicant (Siemens Medical Solutions USA, Inc) — same firm and product code, not necessarily this device · initiated 2025-08-28

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    recall event 97530 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K220151