Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V6.5

K212333

Canon Medical Systems Corporation · cleared 2022-01-24 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.3)
The Diagnostic Ultrasound System Aplio i900 Model TUS-AI900, Aplio i800 Model TUS-AI800, and Aplio i700 Model TUS-AI700 are indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs (thyroid, breast and testicle), trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculo-skeletal (both conventional and superficial), laparoscopic and Thoracic/Pleural. This system provides high-quality ultrasound images in the following modes: B mode, M mode, Continuous Wave, Color Doppler, Pulsed Wave Doppler, Power Doppler and Combination Doppler, as well as Speckle-tracking, Tissue Harmonic Imaging, Combined Modes, Shear wave, Elastography, and Acoustic attenuation mapping. The expansion of marketing language for several previously 510(k) cleared features is also proposed in the subject submission, including the promotion of the support by artificial intelligence (AI) and/or machine learning (ML) of the automated initial contour tracing capability of 2D Wall Motion Tracking for left ventricle (2D WMT) and Auto Ejection Fraction for left ventricle (Auto EF), marketed by Canon as 2D WMT with Full-assist function and Auto EF with Full-assist function, respectively.
Algorithmartificial intelligence (AI) and/or machine learning (ML) of the automated initial contour tracing capability of 2D Wall Motion Tracking for left ventricle (2D WMT) and Auto Ejection Fraction for left ventricle (Auto EF)
source quote (p.6)
The expansion of marketing language for several previously 510(k) cleared features is also proposed in the subject submission, including the promotion of the support by artificial intelligence (AI) and/or machine learning (ML) of the automated initial contour tracing capability of 2D Wall Motion Tracking for left ventricle (2D WMT) and Auto Ejection Fraction for left ventricle (Auto EF), marketed by Canon as 2D WMT with Full-assist function and Auto EF with Full-assist function, respectively.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedYes
source quote (p.7)
Additionally, cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued on October 18, 2018, was included in this submission.

Validation studies (1)

Bench

sample size not stated

standards: 21 CFR § 820, ISO 13485, AAMI/ANSI ES60601-1:2012, IEC 60601-1-2 (2014), IEC 60601-2-37 (2015), IEC 62304 (2015), IEC 62359 (2017), ISO 10993-1 (2009)

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
7
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:97726

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K252074 (decision 2025-10-31) from Canon Medical Systems Corporation for a matching device line ("Aplio i900, Aplio i800 and Aplio i700 Software V9.0 Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K252074

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K241582 (decision 2024-09-12) from Canon Medical Systems Corporation for a matching device line ("Aplio i900/i800/i700 Diagnostic Ultrasound System, Software V7.0 (TUS-AI900, TUS-AI800, TUS-AI700)") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K241582

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K233195 (decision 2024-01-24) from Canon Medical Systems Corporation for a matching device line ("Aplio i900, Aplio i800 and Aplio i700 Software V8.1 Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K233195

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • …and 1 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-09-18

    The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

    recall event 97726 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-03-06

    The battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.

    recall event 96538 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.