Critical Care Suite with Endotracheal Tube Positing AI algorithm

K211161

GE Medical Systems, LLC · cleared 2021-10-29 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
Critical Care Suite is a suite of AI algorithms for the automated image analysis of frontal chest X-rays acquired on a digital x-ray system.
Algorithmdeep learning locked AI algorithm
source quote (p.7)
It is a deep learning locked Al algorithm that can be deployed on several computing platforms such as PACS, On Premise, On Cloud or Imaging Systems.
Adaptive (vs locked)No
source quote (p.6)
nonadaptive machine learning algorithms trained with clinical and/or artificial data.
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

sample size not stated

endpoints: ETT detection; ETT tip to Carina distance measurement; carina localization; ETT tip localization

Reported performance (3 observations)

sensitivity0.9941CI (0.9859, 1.0000)
source quote (p.9)
a sensitivity of 0.9941 (0.9859, 1.0000)
specificity1CI (1.0000, 1.0000)
source quote (p.9)
a specificity of 1.0000 (1.0000, 1.0000)
aurocas written: “auc0.9999CI (0.9998, 1.0000)
source quote (p.9)
an AUC of 0.9999 (0.9998, 1.0000)

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K211161