SIGNA 7.0T

K211118

GE Medical Systems, LLC · cleared 2021-05-13 · product code LNH · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.5)
This 510(k) submission is for the SIGNA 7.0T MR System, and has been triggered by the addition of the AIR Recon DL software feature and inclusion of installed base magnet system upgrades.
AlgorithmAIR Recon DL feature
source quote (p.5)
The software used on the proposed SIGNA 7.0T system has been modified to include the AIR Recon DL feature.
Adaptive (vs locked)No
PCCPNo
Cybersecurity addressedNo

Validation studies (2)

Bench

sample size not stated

endpoints: image quality; SNR; sharpness; low contrast detectability

Reader study (MRMC)

sample size not stated

endpoints: SNR; image sharpness; radiologist preference

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

100
recalls in product code, 24mo
510
MAUDE reports in code, 12mo
+5%
vs code's own 3-yr baseline
8
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253780 (decision 2026-02-06) from Ge Medical Systems, LLC for a matching device line ("SIGNA™ Bolt") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253780

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253779 (decision 2026-02-05) from Ge Medical Systems, LLC for a matching device line ("SIGNA™ Sprint Select") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253779

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K251399 (decision 2025-09-11) from GE Medical Systems, LLC for a matching device line ("SIGNA™ Sprint") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K251399

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K250379 (decision 2025-03-17) from GE Medical Systems, LLC for a matching device line ("SIGNA Prime Elite") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K250379

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K241242 (decision 2024-10-17) from GE Medical Systems, LLC for a matching device line ("SIGNA MAGNUS") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K241242

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K213668 (decision 2022-01-20) from GE Healthcare (GE Medical Systems, LLC) for a matching device line ("SIGNA Hero") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K213668

  • …and 2 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

  • Recalling firm matches this device's applicant (GE Medical Systems, LLC) — same firm and product code, not necessarily this device · initiated 2025-06-20

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    recall event 97187 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K211118