HS40 Diagnostic Ultrasound System

K210426

Samsung Medison CO., LTD. · cleared 2021-05-13 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.3)
The HS40 diagnostic ultrasound system and probes are designed to obtain ultrasound images and analyze body fluids.
AlgorithmFDA source did not state this
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedNo

Validation studies (1)

Bench

sample size not stated

standards: AAMI ANSI ES60601-1:2005/(R)2012, IEC 60601-1:2005, IEC60601-1-2: 2014(4th Edition), IEC 60601-2-37 Edition 2.0 2007, ISO 10993-1:2009/(R)2013, ISO 14971:2007, NEMA UD 2-2004 (R2009)

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
6
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K221599 (decision 2022-08-22) from Samsung Medison Co., Ltd. for a matching device line ("HS40 Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K221599

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K211945 (decision 2021-09-08) from Samsung Medison CO., LTD. for a matching device line ("V8 Diagnostic Ultrasound System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K211945

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97726

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95254

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K210426