K203514

Philips Healthcare (Suzhou) Co., Ltd. · cleared 2021-06-17 · product code JAK · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.5)
Precise Position including a camera with both color and depth function is installed in the ceiling of the scan room, in such a way to cover the entire patient on the patient table. The camera control and image data transmit via the high speed fiber and copper hybrid USB cable. The power supply of the camera is from the gantry. Precise position adopts the AI algorithm (Convolution Neural Network) to detect the joints of the patient body, and then identify surview start/end position and patient orientation. The algorithm can also support detect center of patient anatomy.
AlgorithmAI algorithm (Convolution Neural Network)
source quote (p.5)
Precise position adopts the AI algorithm (Convolution Neural Network) to detect the joints of the patient body, and then identify surview start/end position and patient orientation. The algorithm can also support detect center of patient anatomy.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=80 cases

endpoints: To calculate the time saved per surview planning with and without Precise Position; To calculate the accuracy of vertical (iso)center positioning and Surview (horizontal) start /end position with and without Precise Position; Intra operator consistency in positioning the patient and surview scan range

standards: AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012, IEC 60601-1:2014, ISO 14971, IEC 62304:2015

Reported performance (1 observation)

accuracyas written: “Increase in vertical position accuracy50
source quote (p.9)
with "Precise Position" the vertical position accuracy is increased up to 50%.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

75
recalls in product code, 24mo
192
MAUDE reports in code, 12mo
-22%
vs code's own 3-yr baseline
9
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98738

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98588

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98206

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97699

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97010

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (, initiated 2025-05-23): "Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96967

  • …and 3 more.

Recalls attributed to this device

  • Recalling firm matches this device's applicant (Philips Healthcare (Suzhou) Co., Ltd. Suzhou Industrial Park No.) — same firm and product code, not necessarily this device · initiated 2025-12-03

    Potential for incomplete scan due to unstable connection inside of floating sensor.

    recall event 98238 (openFDA)

  • Recalling firm matches this device's applicant (Philips Healthcare) — same firm and product code, not necessarily this device · initiated 2023-08-02

    Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.

    recall event 93000 (openFDA)

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K203514