Biograph Vision, Biograph mCT Family of PET/CTs

K193248

Siemens Medical Solutions USA, Inc · cleared 2020-02-14 · product code KPS · Radiology

FDA 510(k) summary for Biograph Vision, Biograph mCT Family of PET/CTs (K193248)

Constat organizes this clearance into predicate context, validation evidence, reported performance, postmarket signals, and reimbursement links. Each extracted claim stays tied to the public FDA source; missing fields remain explicitly unstated.

Validation studies
4
Cited predicates
3
Performance observations
1
Product code
KPS

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.3)
“The Siemens Biograph Vision and Biograph mCT PET/CT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. Modifications include, corrections to software anomalies and addition of new software features, including: OncoFreeze AI (Data Driven Gating) FlowMotion AI (PET FAST Planning)”
AlgorithmOncoFreeze AI (Data Driven Gating), FlowMotion AI (PET FAST Planning)
source quote (p.6)
“Modifications include, corrections to software anomalies and addition of new software features, including: OncoFreeze AI (Data Driven Gating) FlowMotion AI (PET FAST Planning)”
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.11)
“Cybersecurity information in accordance with FDA Guidance documents issued October 2, 2014 has been provided. The Biograph Vision and Biograph mCT software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use. Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Biograph Vision, Biograph mCT and external devices.”

Validation studies (4)

Bench

sample size not stated

endpoints: Resolution - Full Size; Count Rate / Scatter / Sensitivity; Image Quality (4 to 1) - (% Contrast / Background Variability); Co-Registration Accuracy

standards: NEMA NU2-2018

Bench

sample size not stated

endpoints: SUVmax; SUVmean; Volume measurement

Bench

sample size not stated

endpoints: Accuracy of bed range definition

Bench

sample size not stated

endpoints: Difference from ground truth in ROI values

Reported performance (1 observation)

accuracyas written: “Co-Registration Accuracy Max Error”stated without value
source quote (p.8)
“Max Error Pass ≤ 5 mm (Biograph Vision, Page 8); Max Error Pass ≤ 5 mm (Biograph mCT, Page 9)”

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

25
recalls in product code, 24mo
8
MAUDE reports in code, 12mo
-72%
vs code's own 3-yr baseline
4
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 9, Andrey Sakharov Haifa Israel, initiated 2025-12-24): "There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most eas" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98269

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel, initiated 2025-06-20): "Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector mo" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97193

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (Hermes Medical Solutions AB Strandbergsgatan 16 Stockholm Sweden, initiated 2024-10-31): "Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95673

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (Canon Medical System, USA, INC., initiated 2024-09-17): "When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95471

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-08-19.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Related FDA clearances

These are recent clearances in the same FDA product-code category. Shared classification does not by itself make a device an appropriate predicate.