7D Surgical System Cranial Biopsy and Ventricular Catheter Placement Application

K192945

7D Surgical Inc. · cleared 2019-11-27 · product code HAW · Neurology

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.3)
The 7D Surgical System is a stereotaxic image guidance system intended for the spatial positioning and orientation of neurosurgical instruments used by surgeons. The system is also intended to be used as the primary surgical luminaire during image guided surgery.
AlgorithmThe Cranial Application software works in conjunction with 7D Surgical Machine Vision Guidance System which consists of clinical software, optically tracked surgical Pointer, a reference frame instrument and platform/computer hardware which is substantially equivalent to K181041. Image guidance tracks the position of instruments in relation to the surgical anatomy and identifies this position on DICOM images or intraoperative structured light images of the patient.
source quote (p.5)
The Cranial Application software works in conjunction with 7D Surgical Machine Vision Guidance System which consists of clinical software, optically tracked surgical Pointer, a reference frame instrument and platform/computer hardware which is substantially equivalent to K181041. Image guidance tracks the position of instruments in relation to the surgical anatomy and identifies this position on DICOM images or intraoperative structured light images of the patient.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Bench

sample size not stated

endpoints: Target Registration Error (TRE); Trajectory Angle Error (ATE)

standards: ISO 10993-1, ISO 17665-1, AAMI / ANSI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-41, IEC 60601-1-6, IEC 60825-1

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

5
recalls in product code, 24mo
1922
MAUDE reports in code, 12mo
-39%
vs code's own 3-yr baseline
1
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code HAW (Howmedica Osteonics Corp., initiated 2025-11-12): "When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an in" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97988

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Neurology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.