SOZO

K190529

ImpediMed Limited · cleared 2019-11-25 · product code QJB · Cardiovascular

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.4)
The SOZO system consists of a connected hand and footplate with built-in stainless steel electrodes, paired with an Android tablet over Bluetooth connection. An app (“SOZOapp”), supplied with the tablet, controls the functionality of the hardware and supplies the bioimpedance measurement data to a database (“SOZOhub") contained within the hospital/facility network.
AlgorithmEstablished algorithms are used to analyze data and calculate various body composition parameters. Basal metabolic rate (BMR; based on Mifflin – St. Jeor's algorithm).
source quote (p.5)
Established algorithms are used to analyze data and calculate various body composition parameters and present the outputs for the clinician to review. Basal metabolic rate (BMR; based on Mifflin – St. Jeor's algorithm) displayed in calories per day
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Bench

sample size not stated

endpoints: verify correlation in outputs

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Cardiovascular panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K190529