Uscan

K181574

EchoNous, Inc. · cleared 2018-07-10 · product code IYO · Radiology

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.7)
The Uscan is a hand-held, diagnostic ultrasound system with an on-screen display. Its purpose is to acquire ultrasound echo data and display it in B-Mode on an off-the-shelf display. Automated bladder volume measurements are supported.
AlgorithmAutomated bladder volume measurements
source quote (p.7)
Automated bladder volume measurements are supported.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedNo

Validation studies (1)

Bench

sample size not stated

endpoints: imaging performance; measurement accuracy

standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62304, IEC 62366, ISO 14971, NEMA UD 2-2004

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

8
recalls in product code, 24mo
344
MAUDE reports in code, 12mo
-40%
vs code's own 3-yr baseline
1
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYO (Civco Medical Instruments Co. Inc., initiated 2026-03-02): "There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98513

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device); a recall is shown as device-attributed only when the recall record itself lists this clearance number. Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to 510(k) AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/K181574