Caption Interpretation Automated Ejection Fraction Software

DEN220063

Caption Health, Inc · granted 2023-02-24 · product code QVD · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.1)
A radiological machine learning based quantitative imaging software with predetermined change control plan is a software-only device which employs machine learning algorithms on radiological images to provide quantitative imaging outputs.
Algorithmmachine learning algorithms, deep learning, convolutional neural network
source quote (p.2)
The Caption Interpretation Automated Ejection Fraction Software ("AutoEF") applies machine learning algorithms to process two-dimensional transthoracic echocardiography images for calculating left ventricular ejection fraction. The software uses an algorithm that was derived through use of deep learning and locked prior to validation. This view selection is based on a convolutional neural network.
Adaptive (vs locked)No
source quote (p.4)
The machine learning algorithms used in Caption Interpretation Auto Ejection Fraction Software are trained, tuned, and "locked" prior to release i.e., they do not continuously learn in the field.
PCCPYes
source quote (p.3)
Caption Interpretation Auto Ejection Fraction Software includes an FDA-authorized predetermined change control plan (PCCP).
Cybersecurity addressedYes
source quote (p.5)
The software documentation included: ... 11. Cybersecurity

Validation studies (2)

Retrospective clinical

n=186 patients · 3 site(s)

endpoints: root mean square deviation (RMSD) between AutoEF derived values and reference standard

Reader study (MRMC)

sample size not stated

endpoints: positive predictive value (PPV); sensitivity

Reported performance (1 observation)

ppvas written: “Positive Predictive Value (PPV)stated without value
source quote (p.5)
That study met the pre-defined acceptance criteria and found that the observed PPV point estimates for the Clip Annotator were greater than 97% for identification of the imaging mode and the view.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: De Novo decision summary PDF.

Predicate network

Postmarket — what happened after clearance

Not yet tracked — the weekly postmarket refresh hasn't snapshotted this device.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

Applicable FDA guidance — what the submission is measured against

FDA guidance documents and guiding principles applicable to De Novo AI/ML devices in the Radiology panel. A curated reference index, not legal or regulatory advice — each item states its own status, and a draft is never binding.

Applicability is derived from the device's FDA advisory panel and pathway — cross-cutting guidances apply to every AI/ML device; panel-specific ones are flagged. Titles, dates, and links verified against fda.gov as of July 2026.

Constat Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: constat.dev/precedent/device/DEN220063